Sputter deposition of titanium on poly(Methyl methacrylate) enhances corneal biocompatibility

Permanent link
https://hdl.handle.net/10037/22894
DOI
https://doi.org/10.1167/TVST.9.13.41
Thumbnail
View/Open
article.pdf (10.32Mb)
Published version (PDF)
Date
2020-12-23
Type
Journal article
Tidsskriftartikkel
Peer reviewed

Author
Sharifi, Sina; Islam, Mohammad Mirazul; Sharifi, Hannah; Islam, Rakibul; Nilsson, Per; Dohlman, Claes H.; Mollnes, Tom Eirik; Paschalis, Eleftherios I.; Chodosh, James
Abstract
Purpose: To evaluate titanium (Ti) sputtering of the poly(methyl methacrylate) (PMMA) stem of the Boston Keratoprosthesis (BK) as a method to enhance interfacial adhesion between the PMMA and the recipient corneal tissue.

Methods: PMMA specimens were plasma treated with Ar/O2 and coated with Ti using a DC magnetron sputtering instrument. The topography and hydrophilicity of the surfaces were characterized using atomic force microscopy and a water contact angle instru ment, respectively. Scratch hardness and adhesion of the Ti film were measured using a mechanical tester. Biocompatibility assessments were performed using cultured human corneal fibroblasts and whole blood ex vivo. The optical quality of the Ti sputtered BK was evaluated using a custom-made optical bench.

Results: By contact angle studies, the Ti coating improved PMMA hydrophilicity to match that of medical-grade Ti (Ti-6Al-4V-ELI). Ti sputtering of contact surfaces resulted in a plate-like morphology with increased surface roughness, without impacting the transparency of the BK optical component. Scratch testing indicated that the mechani cal behavior of the Ti coating was similar to that of casted Ti, and the coating was stable in pull-off adhesion testing. Sputtered Ti film was highly biocompatible based on tests of cell viability, adhesion, proliferation, differentiation, collagen deposition, and kerato can expression, the properties of which exceeded those of uncoated PMMA and did not induce increased complement activation.

Conclusions: Titanium coating of the BK stem generated a mechanically and biolog ically favorable interface, which may help to enhance corneal stromal adhesion and biocompatibility.

Translational Relevance: Improving the biocompatibility of the BK PMMA stem may improve long-term outcomes of implantation.

Publisher
ARVO
Citation
Sharifi, Islam, Sharifi, Islam, Nilsson, Dohlman, Mollnes, Paschalis, Chodosh. Sputter deposition of titanium on poly(Methyl methacrylate) enhances corneal biocompatibility. Translational Vision Science & Technology. 2020;9(13):1-14
Metadata
Show full item record
Collections
  • Artikler, rapporter og annet (klinisk medisin) [1974]
Copyright 2020 The Author(s)