dc.description.abstractThe push broom design of an underwater hyperspectral imaging (UHI) instrument makes it possible to measure angle-resolved spectral radiance L(λ)in a plane. We describe the characterization of a commercial UHI instrument (UHI-4, Ecotone AS, Norway) and the spectral, geometric, and radiometric calibration transfer for measuring L(λ) in absolute units [µW cm<sup>−2</sup> nm<sup>−1</sup> sr<sup>−1</sup> ]. We present a low-cost instrument characterization approach that is intended to be easily replicated for other users to perform their own calibration transfer. Cross-calibration with a RAMSES-ARC spectroradiometer (TriOS, Germany) in air and in water shows good linear correlation across the observed spectral range and over four orders of magnitude in dynamic range, with the UHI instrument providing higher sensitivity overall.en_US
dc.identifier.citationSchartmüller, McKee, Berge, Johnsen. Calibration verification of an underwater hyperspectral imaging push broom instrument to measure light in absolute units and field demonstration. Applied Optics. 2024;63(27):7200-7211en_US
dc.identifier.cristinIDFRIDAID 2300615
dc.identifier.doi10.1364/AO.532003
dc.identifier.issn1559-128X
dc.identifier.issn2155-3165
dc.identifier.urihttps://hdl.handle.net/10037/35055
dc.language.isoengen_US
dc.publisherOptica Publishing Groupen_US
dc.relation.journalApplied Optics
dc.rights.accessRightsopenAccessen_US
dc.rights.holderCopyright 2024 Optica Publishing Groupen_US
dc.rights.urihttps://creativecommons.org/licenses/by/4.0en_US
dc.rightsAttribution 4.0 International (CC BY 4.0)en_US
dc.subjectVDP::Matematikk og naturvitenskap: 400::Fysikk: 430::Elektromagnetisme, akustikk, optikk: 434en_US
dc.subjectVDP::Mathematics and natural scienses: 400::Physics: 430::Electromagnetism, acoustics, optics: 434en_US
dc.subjectVDP::Teknologi: 500::Miljøteknologi: 610en_US
dc.subjectVDP::Technology: 500::Environmental engineering: 610en_US
dc.titleCalibration verification of an underwater hyperspectral imaging push broom instrument to measure light in absolute units and field demonstrationen_US
dc.type.versionpublishedVersionen_US
dc.typeJournal articleen_US
dc.typeTidsskriftartikkelen_US
dc.typePeer revieweden_US


Tilhørende fil(er)

Thumbnail

Denne innførselen finnes i følgende samling(er)

Vis enkel innførsel

Attribution 4.0 International (CC BY 4.0)
Med mindre det står noe annet, er denne innførselens lisens beskrevet som Attribution 4.0 International (CC BY 4.0)