dc.description.abstract | Background: To evaluate repeatability and agreement in measurements of
total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster
700, and MS-39.<p>
<p>Methods: Three consecutive measurements were taken with each device in
136 eyes. TCA values were converted into components J<sub>0</sub> and J<sub>45</sub>. The Anterion
and the IOLMaster 700 also provided axial length (AL) measurements. The
repeatability was calculated using pooled within-subject standard deviation
(S<sub>w</sub>). The agreement among the four devices was assessed by pairwise comparisons and Bland–Altman plots.
<p>Results: For all devices, the repeatability of TCA measurements showed
S<sub>w</sub> ≤0.23 D for TCA magnitude, ≤0.14 D for J<sub>0</sub>, and ≤0.12 D for J<sub>45</sub>. There
were statistically significant differences in TCA magnitude for each pair,
except for IOLMaster 700 with MS-39, and Anterion with MS-39. The repeatability (S<sub>w</sub>) of axis measurements had a statistically significant negative correlation with the TCA magnitude (p < 0.001 for all devices). Both Anterion and
IOLMaster 700 had high repeatability in AL measurements (S<sub>w</sub>: 0.007 mm for
Anterion and 0.009 mm for IOLMaster 700). The difference in AL between the
two was 0.015 ± 0.033 mm (p < 0.001).
<p>Conclusions: All four devices showed good repeatability in TCA measurements in keratoconic eyes, the agreement for TCA measurements between the
tested devices was generally low. Anterion and IOLMaster 700 showed good
repeatability and agreement in AL measurements. | en_US |
dc.identifier.citation | Feng, Nitter, Bertelsen, Stojanovic. Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices. Clinical and Experimental Ophthalmology (CEO). 2024;52(8):800-810 | en_US |
dc.identifier.cristinID | FRIDAID 2348364 | |
dc.identifier.doi | 10.1111/ceo.14423 | |
dc.identifier.issn | 1442-6404 | |
dc.identifier.issn | 1442-9071 | |
dc.identifier.uri | https://hdl.handle.net/10037/36404 | |
dc.language.iso | eng | en_US |
dc.publisher | Wiley | en_US |
dc.relation.journal | Clinical and Experimental Ophthalmology (CEO) | |
dc.rights.accessRights | openAccess | en_US |
dc.rights.holder | Copyright 2024 The Author(s) | en_US |
dc.rights.uri | https://creativecommons.org/licenses/by-nc-nd/4.0 | en_US |
dc.rights | Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) | en_US |
dc.title | Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices | en_US |