dc.description.abstract | People with type 1 diabetes and impaired awareness of hypoglycaemia (IAH) are prone to severe
hypoglycaemia. Previous attempts to develop non-invasive hypoglycaemia alarm systems have shown
promising results, but it is not known if such alarms can detect severe hypoglycaemia in people with
IAH. We aimed to explore whether a combination of non-invasive sensors could reliably evaluate
hypoglycaemia (plasma glucose (PG) minimum 2.5mmol/L) in people with IAH. Twenty participants
with type 1 diabetes and IAH underwent randomly ordered, single blinded hyperinsulinemic
euglycaemic and hyperinsulinemic hypoglycaemic clamps. Sweating, skin temperature, ECG,
counterregulatory hormones and symptoms of hypoglycaemia were assessed. Overall, we were not
able to detect clamp-induced hypoglycaemia with sufcient sensitivity and specifcity for further
clinical use. As a post-hoc analysis, we stratifed participants according to their ability to identify
hypoglycaemic symptoms during hypoglycaemic clamps. Five out of 20 participants could identify
such symptoms. These participants had a signifcantly higher adrenaline response to hypoglycaemia
(p<0.001) and were reliably identifed by sensors. Based on our observations, a non-invasive alarm
system based on measurement of sweating responses and ECG changes during hypoglycaemia might
provide an alert at a plasma glucose concentration around 2.5mmol/L if an adequate sympatho-adrenal
reaction is elicited. | en_US |
dc.description | Source at: <a href=http://doi.org/10.1038/s41598-018-33189-1> http://doi.org/10.1038/s41598-018-33189-1</a> | en_US |
dc.identifier.citation | Elvebakk, O., Tronstad, C., Birkeland, K. I., Jenssen, T. G., Bjørgaas, M. R. R., Frøslie, K. F., ... Gulseth, H. L. (2018). Evaluation of hypoglycaemia with non-invasive sensors in people with type 1 diabetes and impaired awareness of hypoglycaemia. Scientific Reports, 8, 14722, 1-11. http://doi.org/10.1038/s41598-018-33189-1 | en_US |
dc.identifier.cristinID | FRIDAID 1623045 | |
dc.identifier.doi | 10.1038/s41598-018-33189-1 | |
dc.identifier.issn | 2045-2322 | |
dc.identifier.uri | https://hdl.handle.net/10037/14059 | |
dc.language.iso | eng | en_US |
dc.publisher | Nature Research | en_US |
dc.relation.journal | Scientific Reports | |
dc.rights.accessRights | openAccess | en_US |
dc.subject | VDP::Medical disciplines: 700::Clinical medical disciplines: 750 | en_US |
dc.subject | VDP::Medisinske Fag: 700::Klinisk medisinske fag: 750 | en_US |
dc.subject | VDP::Medical disciplines: 700::Clinical medical disciplines: 750::Hematology: 775 | en_US |
dc.subject | VDP::Medisinske Fag: 700::Klinisk medisinske fag: 750::Hematologi: 775 | en_US |
dc.subject | VDP::Medical disciplines: 700::Clinical medical disciplines: 750::General internal medicine: 770 | en_US |
dc.subject | VDP::Medisinske Fag: 700::Klinisk medisinske fag: 750::Generell indremedisin: 770 | en_US |
dc.title | Evaluation of hypoglycaemia with non-invasive sensors in people with type 1 diabetes and impaired awareness of hypoglycaemia | en_US |
dc.type | Journal article | en_US |
dc.type | Tidsskriftartikkel | en_US |
dc.type | Peer reviewed | en_US |