Electrosurgery and temperature increase in tissue with a passive metal implant
Permanent lenke
https://hdl.handle.net/10037/17485Dato
2019-03-12Type
Journal articleTidsskriftartikkel
Peer reviewed
Forfatter
Martinsen, Tormod; Pettersen, Fred-Johan; Kalvøy, Håvard; Tronstad, Christian; Kvarstein, Gunnvald; Bakken, André; Høgetveit, Jan Olav; Martinsen, Ørjan Grøttem; Grimnes, Sverre; Frich, Lars IngeSammendrag
Objective: To determine the temperature of tissue close to metal implants during electrosurgery in an in-vitro model.
Design, Setting, and Participants: Thirty tissue samples (15 with a metal implant placed in center, 15 controls without implant) were placed in an in vitro measurement chamber. Electrosurgery was applied at 5–60 W with the active electrode at three defined distances from the implant while temperatures at four defined distances from the implant were measured using fiber-optic sensors.
Main Outcomes and Measures: Tissue temperature increase at the four tissue sites was determined for all power levels and each of the electrode-to-implant distances. Based on a linear mixed effects model analysis, the primary outcomes were the difference in temperature increase between implant and control tissue, and the estimated temperature increase per watt per minute.
Results: Tissues with an implant had higher temperature increases than controls at all power levels after 1 min of applied electrosurgery (mean difference of 0.16°C at 5 W, 0.50°C at 15 W, 1.11°C at 30 W, and 2.22°C at 60 W, all with p < 0.001). Temperature increase close to the implant was estimated to be 0.088°C/W/min (95% CI: 0.078–0.099°C/W/min; p < 0.001). Temperature could increase to above 43°C after 1 min of 60 W. Active electrode position had no significant effect on temperature increases for tissues with implant (p = 0.6).
Conclusions and Relevance: The temperature of tissue close to a metal implant increases with passing electrosurgery current. There is a significant risk of high tissue temperature when long activation times or high power levels are used.