Sputter deposition of titanium on poly(Methyl methacrylate) enhances corneal biocompatibility
Permanent lenke
https://hdl.handle.net/10037/22894Dato
2020-12-23Type
Journal articleTidsskriftartikkel
Peer reviewed
Forfatter
Sharifi, Sina; Islam, Mohammad Mirazul; Sharifi, Hannah; Islam, Rakibul; Nilsson, Per; Dohlman, Claes H.; Mollnes, Tom Eirik; Paschalis, Eleftherios I.; Chodosh, JamesSammendrag
Methods: PMMA specimens were plasma treated with Ar/O2 and coated with Ti using a DC magnetron sputtering instrument. The topography and hydrophilicity of the surfaces were characterized using atomic force microscopy and a water contact angle instru ment, respectively. Scratch hardness and adhesion of the Ti film were measured using a mechanical tester. Biocompatibility assessments were performed using cultured human corneal fibroblasts and whole blood ex vivo. The optical quality of the Ti sputtered BK was evaluated using a custom-made optical bench.
Results: By contact angle studies, the Ti coating improved PMMA hydrophilicity to match that of medical-grade Ti (Ti-6Al-4V-ELI). Ti sputtering of contact surfaces resulted in a plate-like morphology with increased surface roughness, without impacting the transparency of the BK optical component. Scratch testing indicated that the mechani cal behavior of the Ti coating was similar to that of casted Ti, and the coating was stable in pull-off adhesion testing. Sputtered Ti film was highly biocompatible based on tests of cell viability, adhesion, proliferation, differentiation, collagen deposition, and kerato can expression, the properties of which exceeded those of uncoated PMMA and did not induce increased complement activation.
Conclusions: Titanium coating of the BK stem generated a mechanically and biolog ically favorable interface, which may help to enhance corneal stromal adhesion and biocompatibility.
Translational Relevance: Improving the biocompatibility of the BK PMMA stem may improve long-term outcomes of implantation.